LABPROOF AI

Operations

DashboardRecordsUpload

Review

Violations41Reviews21

Reporting

ReportsAudit trailAnalytics

Workspace

Settings
Audit readiness
92%Q2 ready
All records

ENV-2026-04403

FlaggedEnvironmental

Groundwater Monitoring Well MW-7

Route for review Add to report
82Integrity

2

Findings

1

Critical

1

Cleared

Method
EPA 8260D — Volatile Organics (GC/MS)
Instrument
Agilent 7900 ICP-MS
Analyst
K. Abara
Sample ID
ENV-S19561
Batch
B-2295
Source
scanned_coc_packet.pdf
Collected
Jun 17, 2026
Scanned
Jun 27, 2026

ALCOA+ findings (2)

Confidence96%Fieldentry_timestamp AI-assisted · human sign-off required
Why flagged
The record's entry timestamp precedes the instrument acquisition time, indicating data was entered out of sequence. Contemporaneous capture requires recording at the time work is performed.
Source evidence
entry_timestamp 09:02 < acquisition_time 11:47 on the same worksheet.
Suggested remediation
Confirm the true acquisition time from raw instrument metadata and correct the entry timestamp with a documented justification.
Reviewer decision
Awaiting a reviewer. No decision recorded yet.
Record history

Imported from scanned_coc_packet.pdf

Jun 17, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 27, 2026, 02:20 PM

Backdated entry detected flagged (critical)

Jun 27, 2026, 02:20 PM

Timezone inconsistency in timestamps flagged (minor)

Jun 27, 2026, 02:20 PM