LABPROOF AI

Operations

DashboardRecordsUpload

Review

Violations41Reviews21

Reporting

ReportsAudit trailAnalytics

Workspace

Settings
Audit readiness
92%Q2 ready
All records

ENV-2026-04405

In reviewEnvironmental

Groundwater Monitoring Well MW-7

Route for review Add to report
74Integrity

3

Findings

1

Critical

0

Cleared

Method
EPA 8260D — Volatile Organics (GC/MS)
Instrument
Waters Xevo TQ-S micro
Analyst
N. Okafor
Sample ID
ENV-S34581
Batch
B-2266
Source
manual_entry
Collected
Jun 3, 2026
Scanned
Jun 24, 2026

ALCOA+ findings (3)

Confidence94%Fieldintegration_mode AI-assisted · human sign-off required
Why flagged
Chromatographic peaks were manually integrated but no justification or before/after comparison was retained. Manual integration must be documented to remain defensible.
Source evidence
integration_mode = manual on 3 peaks; justification_note empty.
Suggested remediation
Attach the auto vs. manual integration comparison and a signed justification per the data-integrity SOP.
Reviewer decision
Awaiting a reviewer. No decision recorded yet.
Record history

Imported from manual_entry

Jun 3, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 24, 2026, 02:20 PM

Manual integration without justification flagged (major)

Jun 24, 2026, 02:20 PM

Timezone inconsistency in timestamps flagged (minor)

Jun 24, 2026, 02:20 PM

Result modified after approval without reason flagged (critical)

Jun 24, 2026, 02:20 PM