LABPROOF AI

Operations

DashboardRecordsUpload

Review

Violations41Reviews21

Reporting

ReportsAudit trailAnalytics

Workspace

Settings
Audit readiness
92%Q2 ready
All records

ENV-2026-04406

In reviewEnvironmental

Groundwater Monitoring Well MW-7

Route for review Add to report
63Integrity

4

Findings

1

Critical

1

Cleared

Method
EPA 6020B — Metals by ICP-MS
Instrument
Sciex Triple Quad 6500+
Analyst
T. Morales
Sample ID
ENV-S73734
Batch
B-2276
Source
manual_entry
Collected
Jun 17, 2026
Scanned
Jun 27, 2026

ALCOA+ findings (4)

Confidence96%Fieldraw_data_ref AI-assisted · human sign-off required
Why flagged
The reported result is not linked to its source raw-data file, so the underlying evidence cannot be retrieved on demand. Available records must be reproducible from source.
Source evidence
raw_data_ref points to a path that no longer resolves on the instrument share.
Suggested remediation
Link the instrument raw-data export to the record and verify the file opens with the native acquisition software.
Reviewer decision
Yuki Nakamura (Reviewer) marked this in review.
Record history

Imported from manual_entry

Jun 17, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 27, 2026, 02:20 PM

Raw data file not linked flagged (major)

Jun 27, 2026, 02:20 PM

Manual integration without justification flagged (major)

Jun 27, 2026, 02:20 PM

Calibration record incomplete flagged (major)

Jun 27, 2026, 02:20 PM