ENV-2026-04411

FlaggedEnvironmental

Drinking Water — Distribution Point D-12

Route for review Add to report
65Integrity

3

Findings

2

Critical

1

Cleared

Method
EPA 6020B — Metals by ICP-MS
Instrument
Waters Xevo TQ-S micro
Analyst
K. Abara
Sample ID
ENV-S18715
Batch
B-2233
Source
bench_worksheet_Q2.xlsx
Collected
Jun 18, 2026
Scanned
Jun 24, 2026

ALCOA+ findings (3)

Confidence98%Fieldintegration_mode AI-assisted · human sign-off required
Why flagged
Chromatographic peaks were manually integrated but no justification or before/after comparison was retained. Manual integration must be documented to remain defensible.
Source evidence
integration_mode = manual on 3 peaks; justification_note empty.
Suggested remediation
Attach the auto vs. manual integration comparison and a signed justification per the data-integrity SOP.
Reviewer decision
Rafael Silva (Reviewer) marked this in review.
Record history

Imported from bench_worksheet_Q2.xlsx

Jun 18, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 24, 2026, 02:20 PM

Manual integration without justification flagged (major)

Jun 24, 2026, 02:20 PM

Result modified after approval without reason flagged (critical)

Jun 24, 2026, 02:20 PM

Transcription differs from instrument export flagged (critical)