FOD-2026-04412

In reviewFood

Spice Blend — Mycotoxin Check

Route for review Add to report
68Integrity

2

Findings

2

Critical

0

Cleared

Method
ISO 6579 — Salmonella Detection
Instrument
Agilent 8890 GC
Analyst
R. Cole
Sample ID
FOD-S98345
Batch
B-2208
Source
scanned_coc_packet.pdf
Collected
Jun 2, 2026
Scanned
Jun 24, 2026

ALCOA+ findings (2)

Confidence83%Fieldreported_value AI-assisted · human sign-off required
Why flagged
A reported value was edited after it had moved to an approved state, and no change reason was captured. Silent edits to original data break the chain of custody.
Source evidence
Value changed 18.4 → 12.1 µg/L; no entry in change_reason; edit timestamp post-dates approval.
Suggested remediation
Revert to the original instrument value or attach a signed change-control note documenting the reason, prior value, and authorizing reviewer.
Reviewer decision
Awaiting a reviewer. No decision recorded yet.
Record history

Imported from scanned_coc_packet.pdf

Jun 2, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 24, 2026, 02:20 PM

Result modified after approval without reason flagged (critical)

Jun 24, 2026, 02:20 PM

Result modified after approval without reason flagged (critical)