CAN-2026-04434

In reviewCannabis

Vape Cartridge Heavy Metals

Route for review Add to report
91Integrity

1

Findings

0

Critical

0

Cleared

Method
Microbial Contaminants (qPCR)
Instrument
Waters Xevo TQ-S micro
Analyst
M. Delgado
Sample ID
CAN-S30697
Batch
B-2205
Source
LIMS_export_2026-06.csv
Collected
Jun 7, 2026
Scanned
Jun 23, 2026

ALCOA+ findings (1)

Confidence85%Fieldintegration_mode AI-assisted · human sign-off required
Why flagged
Chromatographic peaks were manually integrated but no justification or before/after comparison was retained. Manual integration must be documented to remain defensible.
Source evidence
integration_mode = manual on 3 peaks; justification_note empty.
Suggested remediation
Attach the auto vs. manual integration comparison and a signed justification per the data-integrity SOP.
Reviewer decision
Awaiting a reviewer. No decision recorded yet.
Record history

Imported from LIMS_export_2026-06.csv

Jun 7, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 23, 2026, 02:20 PM

Manual integration without justification flagged (major)