All records

ENV-2026-04450

In reviewEnvironmental

Groundwater Monitoring Well MW-7

Route for review Add to report
61Integrity

3

Findings

2

Critical

1

Cleared

Method
EPA 6020B — Metals by ICP-MS
Instrument
Agilent 8890 GC
Analyst
A. Fontaine
Sample ID
ENV-S83910
Batch
B-2224
Source
manual_entry
Collected
Jun 16, 2026
Scanned
Jun 22, 2026

ALCOA+ findings (3)

Confidence83%Fieldsample_id AI-assisted · human sign-off required
Why flagged
The sample identifier on the bench worksheet does not match the identifier in the instrument sequence. Inconsistent identifiers create ambiguity about which sample produced the result.
Source evidence
Worksheet shows ENV-2026-04412; sequence file lists ENV-2026-04421.
Suggested remediation
Reconcile the identifiers against the chain-of-custody form and correct the mismatched field with a documented note.
Reviewer decision
Daniel Chen (Reviewer) marked this approved.
Decision recorded · approved
Record history

Imported from manual_entry

Jun 16, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 22, 2026, 02:20 PM

Sample ID mismatch across worksheet flagged (major)

Jun 22, 2026, 02:20 PM

Transcription differs from instrument export flagged (critical)

Jun 22, 2026, 02:20 PM

Transcription differs from instrument export flagged (critical)

Jun 22, 2026, 02:20 PM