All records

COS-2026-04453

ApprovedCosmetics

Mineral Foundation Batch MF-0488

Route for review Add to report
89Integrity

1

Findings

0

Critical

0

Cleared

Method
USP <61> — Microbial Enumeration
Instrument
Waters Xevo TQ-S micro
Analyst
R. Cole
Sample ID
COS-S36838
Batch
B-2258
Source
instrument_run_seq.csv
Collected
Jun 22, 2026
Scanned
Jun 25, 2026

ALCOA+ findings (1)

Confidence98%Fieldintegration_mode AI-assisted · human sign-off required
Why flagged
Chromatographic peaks were manually integrated but no justification or before/after comparison was retained. Manual integration must be documented to remain defensible.
Source evidence
integration_mode = manual on 3 peaks; justification_note empty.
Suggested remediation
Attach the auto vs. manual integration comparison and a signed justification per the data-integrity SOP.
Reviewer decision
Awaiting a reviewer. No decision recorded yet.
Record history

Imported from instrument_run_seq.csv

Jun 22, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 25, 2026, 02:20 PM

Manual integration without justification flagged (major)

Jun 25, 2026, 02:20 PM