LABPROOF AI

Operations

DashboardRecordsUpload

Review

Violations41Reviews21

Reporting

ReportsAudit trailAnalytics

Workspace

Settings
Audit readiness
92%Q2 ready
All records

ENV-2026-04458

In reviewEnvironmental

Groundwater Monitoring Well MW-7

Route for review Add to report
64Integrity

3

Findings

2

Critical

1

Cleared

Method
EPA 6020B — Metals by ICP-MS
Instrument
Agilent 7900 ICP-MS
Analyst
L. Bianchi
Sample ID
ENV-S34378
Batch
B-2246
Source
scanned_coc_packet.pdf
Collected
Jun 25, 2026
Scanned
Jun 27, 2026

ALCOA+ findings (3)

Confidence99%Fieldspec_flag AI-assisted · human sign-off required
Why flagged
A measured value exceeds the method's acceptance limit but was not flagged as out-of-specification. Unflagged OOS results can be reported as passing in error.
Source evidence
Lead = 11.8 µg/L against action limit 10.0 µg/L; spec_flag = none.
Suggested remediation
Apply the OOS flag, initiate an investigation per SOP-QA-014, and route the record for reviewer confirmation before reporting.
Reviewer decision
Awaiting a reviewer. No decision recorded yet.
Record history

Imported from scanned_coc_packet.pdf

Jun 25, 2026, 02:20 PM

ALCOA+ integrity scan completed

Jun 27, 2026, 02:20 PM

Out-of-specification value not flagged flagged (major)

Jun 27, 2026, 02:20 PM

Transcription differs from instrument export flagged (critical)

Jun 27, 2026, 02:20 PM

Backdated entry detected flagged (critical)

Jun 27, 2026, 02:20 PM